ciA Standards
:
Ds301 Version 4.2 & DS401 Version 3.0
ISO standards
:
ISO11898 (+ Transceiver TJA1050)
PDO
:
siehe Dokumente
DS-CAN-01
CANopen pressure transmitter with CAN bus signal output
Versatile, media-compatible pressure transmitter with CAN bus interface.
technical properties
Pressure type
:
Relativdruck gegen Umgebungsdruck bzw. gegen interne Atmosphäre
measuring resolution
:
10 Bit
Measuring accuracy
:
Klasse 0,5 bei RT
Total error
:
< 1,5 % v.E bei -10 °C … +80 °C.
measuring grid
:
ab 5 ms
CAN protocol
:
CANopen 2.0 A
Physical Layer
:
nach DIN 11898
ambient conditions
Operating temperature range
:
-40 °C … +80 °C
Storage temperature range
:
-40 °C … +120 °C
Shock resistance
:
30 g, Dauer
Vibration resistance
:
10 g bei 20-1000 Hz
EMC/ESD
:
EN 50082-1 und EN 50082-2
Duration
:
14 ms bei RT
Pressure measuring ranges
kPa
:
bar
0...500
:
0…5
0...1.000
:
0…10
0...2.000
:
0…20
0...5.000
:
0…50
0...10.000
:
0…100
0...35.000
:
0…350
0...50.000
:
0…500
0...70.000
:
0…700
0...200.000
:
0…2.000
0...400.000
:
0…4.000
Standard pressure range
:
in kPa
Non-standard ranges
:
auf Anfrage erhältlich
TK-101
Measuring transducer (R/I-transducer)
The measuring transducer Tk101 and Tk101Ex are built directly into the connection boxes of our Liquid Level Measuring Instruments of type TK-30.
Properties
Operating temperature
:
-20...+60 °C
Storage temperature
:
-40...+80 °C
Ex-protection
:
siehe Zertifikat
CE marking
:
siehe Konformitätserklärung
Input
Resistance range in three-wire circuit
:
4...10 kΩ
Measuring current
:
0,8 mA @ RE = 4 kΩ
Measuring current
:
0,33 mA bei RE = 10 kΩ
Measuring voltage
:
3,3 V
Supply/Output
2-wire
:
4...20 mA
Voltage(Um)
:
11...28 V DC
Burden / load
:
0...800 Ω |RBürde max = [(Um - Um min) / 0,02] Ω
Output signal in case of error
:
max. 24 mA
Fault influences
Max. Total error
:
± 1 %
Temperature influence
:
-0,015 % / °C
Supply voltage influence
:
0,025 % / V
Setting range
Full deflection
:
60...105 %
Zero point adjustment
:
0...25 %
XR-6x0 | XR-6x1
Contact protection relay Measuring transducer
The contact protection relays are universal measuring transducers, which
input signals (resistance values) with "Namur" specification (DIN EN 60947).
Namur" specification (DIN EN 60947) into switching signals of floating
output contacts.
Protection class EN60529
clamp
:
IP20
Housing
:
IP40
Technical Data
Operating temperature
:
-20...+60 °C
Storage temperature
:
-30...+80 °C
Output contacts
:
potentialfrei
Switching voltage
:
min. 5 V, max. 250 V AC / 150 V DC
Switching current
:
min. 5 mA, max. 5 A AC / 5 A DC
Switching capacity
:
min. 300 mW, max.100 VA / max. 50 W